Paradigm ambulatory insulin infusion pumps - risk of compromised insulin therapy

(Medtronic) May lead to severe hypoglycaemia (low blood glucose) or severe hyperglycaemia (high blood glucose). (MDA/2013/035)

CAS deadlines

Action underway: 11 June 2013, action complete: 2 July 2013

Note: These deadlines are for staff and users to be aware of the problem and the advice as recommended by the manufacturer in the FSN.

Device

Medical Device Alert: Paradigm ambulatory insulin infusion pumps manufactured by Medtronic

Image of Paradigm ambulatory insulin infusion pumps.

Paradigm ambulatory insulin infusion pumps.

Manufactured by Medtronic.

Models: MMT - 511, 512, 712, 712E, 515, 715, 522, 722, 722K, 523, 723, 723K, 554 (VEO) and 754 (VEO).

Problem

Risk of compromised insulin therapy, which may lead to severe hypoglycaemia (low blood glucose) or severe hyperglycaemia (high blood glucose). The root causes are:

  • Loose drive support cap (all models)
  • Damage to pump if immersed in water (all models)
  • Programming issue related to the sensor graph timeout (Paradigm VEO MMT-554 and MMT-754 only).
  1. Loose drive support cap:
    Medtronic has received a report of a serious injury related to a loose drive support cap. This may become detached from the pump case, due to impact damage. An attempt to press back the drive support cap may result in an unintended delivery of insulin.

  2. Water damage:
    This may result in a pump alarm or may cause the buttons to stop working.

  3. Sensor graph timeout:
    This issue only applies to Paradigm VEO (MMT-554 and MMT-754) pump users who are also using both Medtronic ‘Continuous Glucose Monitoring’ and the ‘Low Glucose Suspend’ feature.

Paradigm VEO is equipped with a ’Low Glucose Suspend’ feature, which will automatically suspend insulin delivery when glucose levels are too low, to help reduce the risk of severe hypoglycaemia. Basal insulin delivery may not resume in time if the pump user has the ‘Low Glucose Suspend’ feature enabled and has changed the sensor graph timeout setting to “NONE” from the default setting of two minutes. This could result in hyperglycaemia.

Action

  • Identify affected devices.
  • Ensure that all staff and patients receive a copy of Medtronic’s Field Safety Notice (FSN) dated 28 March 2013 and are aware of the recommendations. In particular:
    • Discontinue pump use if drive support cap is loose and contact the manufacturer.
    • Do not immerse the pump in water.
    • Regularly examine the pump for damage, especially after a shock or drop.
    • For Paradigm VEO models, check that the default setting for the sensor graph timeout is not ‘NONE’, if the ‘Low Glucose Suspend’ feature is used.

Action by

  • Diabetes departments
  • Those responsible for the use, service and maintenance of these devices.

Distribution

This MDA has been sent to:  

  • Care Quality Commission (CQC) (headquarters) for information
  • Clinical commissioning groups (CCGs)
  • HSC trusts in Northern Ireland (chief executives)
  • Local authorities in Scotland (equipment coordinators)
  • NHS boards and trusts in Wales (chief executives)
  • NHS boards in Scotland (equipment coordinators)
  • NHS England area teams
  • NHS England area teams (chief executives)
  • NHS England regional teams
  • NHS trusts in England (chief executives)

Onward distribution

Please bring this notice to the attention of relevant employees in your establishment.  Below is a suggested list of recipients.  

Trusts
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:

  • Clinical governance leads
  • Community children’s nurses
  • Community diabetes specialist nurses
  • Community hospitals
  • Diabetes clinics/outpatients
  • Diabetes nurse specialists
  • Diabetes, directors of
  • Diabetologists
  • EBME departments
  • Equipment stores
  • Hospital pharmacies
  • Medical directors
  • Medical libraries
  • Nursing executive directors
  • Paediatric nurse specialists
  • Paediatricians
  • Risk managers
  • Supplies managers

NHS England local area teams
CAS liaison officers for onward distribution to all relevant staff including:

  • Community pharmacists
  • General practitioners

Independent distribution

Establishments registered with the Care Quality Commission (CQC) (England only)
This alert should be read by:

  • Clinics
  • Hospitals in the independent sector
  • Independent treatment centres
  • Private medical practitioners

Please note: CQC and OFSTED do not distribute these alerts. Independent healthcare providers and social care providers can sign up to receive MDAs directly from the Department of Health’s Central Alerting System (CAS) by sending an email to: [email protected] and requesting this facility.

Manufacturer contact

Lezlie Bridge

Regulatory Affairs Manager, UK & Ireland

Medtronic Ltd

Building 9,
Croxley Green Business Park

Watford, WD18 8WW

Tel: 01923 212 213

Fax: 01923 241 004

Email: [email protected]

Feedback

If you have any comments or feedback on this Medical Device Alert please email us at:[email protected]

England

If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2013/035 or 2013/003/027/291/018.

Technical aspects 

Enitan Taiwo or Sharon Knight

Medicines & Healthcare products Regulatory Agency

Floor 4,
151 Buckingham Palace Road

London SW1W 9SZ


Tel: 020 3080 7122 / 7202

Fax: 020 8754 3965

Email: [email protected] or [email protected]

Clinical aspects 

Mark Grumbridge

Medicines & Healthcare products Regulatory Agency

Floor 4,
151 Buckingham Palace Road

London SW1W 9SZ

Tel: 020 3080 7128

Fax: 020 8754 3965

Email: [email protected]

How to report adverse incidents

Please report via our website: Reporting adverse incidents involving medical devices

Further information about CAS can be found on the CAS website

Northern Ireland

Alerts in Northern Ireland will continue to be distributed via the NI SABS system.

Enquiries and adverse incident reports in Northern Ireland should be addressed to:

Northern Ireland Adverse Incident Centre

Health Estates Investment Group

Room 17

Annex 6

Castle Buildings

Stormont Estate

Dundonald BT4 3SQ

Tel: 02890 523 704

Fax: 02890 523 900

Email: [email protected]

How to report adverse incidents in Northern Ireland

Please report directly to NIAIC, further information can be found on the NIAIC website

Further information about SABS can be found on the SABS website

Scotland

Enquiries and adverse incident reports in Scotland should be addressed to:

Incident Reporting and Investigation Centre

Health Facilities Scotland

NHS National Services Scotland

Gyle Square

1 South Gyle Crescent

Edinburgh EH12 9EB

Tel: 0131 275 7575

Fax: 0131 314 0722

Email: [email protected]

Health Facilities Scotland Incident Reporting and Investigation Centre (IRIC)

Wales

Enquiries in Wales should be addressed to:

Improving Patient Safety Team

Medical Directorate

Welsh Government

Cathays Park

Cardiff CF10 3NQ

Tel: 029 2082 3922

Email: [email protected]

Download documents

Medical Device Alert: Paradigm ambulatory insulin infusion pumps manufactured by Medtronic (MDA/2013/035) (71Kb)

Updates to this page

Published 17 December 2014